Veronics Instruments Inc. represents the GE Industrial Sensing’s (Formerly Panametrics and General Eastern) line of reliable Hygrometers and Moisture Meters with a combination of features to meet your needs and budget. Learn more about the following products:

GE’s Aurora analyzer uses tunable diode laser absorption spectroscopy (TDLAS) to measure moisture in natural gas rapidly and accurately. The analyzer is suitable for installation in hazardous areas and operates over a wide range of environmental conditions. Aurora’s fast response quickly alerts and documents when moisture concentrations are out of compliance. Once process upsets are corrected and the gas dries out, the fast response enables natural gas to be cleared for entry into the “energy grid.”

GE’s experience in moisture measurement includes aluminum oxide, polymer capacitance and chilled mirror sensors. GE’s heritage also includes the development of the first injection diode laser in 1962 by Dr. Robert Hall in Schenectady, NY. The compact and inexpensive diode laser made it possible for compact discs, laser printers and fiber optic telecommunications. The new Aurora TDLAS hygrometer enables natural gas processing and transportation facilities to monitor moisture content in real time with high precision and reliability.

Power

Analyzer 100-240 VAC, 50-60 Hz, 10 W
18-32 VDC (24 VDC nominal), 10 W
Optional Electrical Heater 120 VAC, 120 W
230 VAC, 75 W

Range

Parts Per Million by Volume 5 to 5000 ppm
Dew/Frost Point1 -65.5° to -2.6°C frost point-85.9° to 27.3°F frost point
Process Dew/Frost Point1 Process or equivalent dew point/frost point by calculation with process pressure signal (4-20 mA) or constant
Absolute Humidity 3.8 to 3,803 mg/m3
Absolute Humidity 0.24 to 237 lbs/MMSCF
1 Readings below 0°C (32°F) are in “frost point,” temperature and above 0°C (32°F) are in “dew point” temperature.

Accuracy

Parts Per Million by Volume ±2% of reading or 4 ppmv
(Accuracy of other parameters derived from ppmv.)

Response Time

Optical Response <2 seconds
System Response The system response is dependent on the length of sample tubing, sample system components, flow rate and pressure, as well as the change in moisture concentration.

Operating Pressure

Operating Sample Cell Pressure 69 to 172 Kpa (10-25 psia)
Maximum Pressure 1380 KPa (200 psi)
Process Pressure 10,342 KPa (1500 psig) maximum2
2 Higher pressure available with application of additional sampling system components.

Temperature

Operating -20 to 65°C (-4 to 149°F)
Storage -20 to 70°C (-4 to 158°F)
Optional Heater/Thermostat Set Point 25°C (77°F)

Flowrate

Sample Cell Flowrate 10 to 60 SLH (0.4 to 2 SCFH ); 30 SLH (1 SCFH) nominal
Coalescer By-pass Fast Loop 5 to 10X of flowrate through sample cell

Display

Backlit LCD. Three programable simultaneous parameters. Alphanumeric status and diagnostic display. LEDs for power, laser temperature stability, keypad lockout.

I/O

Analog Outputs Three programmable 0/4-20 mA; 500 Ohm max load.
Analog Input Loop powered 4-20 mA input for remote pressure transmitter. Aurora supplies 24 VDC.
Digital Interface Two programmable digital communications ports RS232, RS485 with multidrop capability and assignable address, Modbus RTU protocol

Enclosure

Laser

Class 1 product. Conforms to IEC 60825-1. Edition 2.0 Safety of Laser Products

User Interface

Programmable “through-the-glass“ via magnetic stylus

Net Weight

37 Kg (100 lb)

Hazardous Area Certification

USA/Canada Explosion-proof for Class I, Division 1, Groups C&D
EU and Elsewhere ATEX and IEC Ex:
Ex de IIB T6 -20°C to +65°C
Flameproof with increased safety compartment

Certification

CE Complies with EMC Directive 2004/108/EC, Low Voltage Directive 2006/95/EC and Pressure Directive 97/25/EC for DN/25

Recommended Factory Verification/Service

Five years

Moisture monitoring in oil & gas and transportation facilities.

Moisture Image Series 1 (MIS 1)

The Moisture Image Series 1 (MIS 1) analyzer is the top-of-the-line model in GE Panametrics’ family of Moisture Series analyzers. This multifunction analyzer uses charts, graphs and windows to display any of six parameters (moisture, oxygen, temperature, pressure and two auxiliary inputs) for as many as six channels simultaneously on a large, backlit LCD. Real-time data can also be exported to a PC using the instrument’s RS232 port and standard or custom software packages.

Electronics

Intrinsic Safety Galvanic isolation and energy-limiting circuitry for all inputs (except auxiliary):
BAS01ATEX7097 II (1) G [EEx ia] IIC –20°C to +50°C;
CSA C US Class I, Division 2, Groups B,C&D T4A–Assoc.
Elec. Apparatus [Exia] and Class I, Division 1,Groups A,B,C&D T4
European Compliance Complies with EMC Directive 89/336/EEC and 73/23/EEC LVD (Installation Category II, Pollution Degree 2)
Inputs
  • Moisture: One to six channels
  • Temperature: One to six channels
  • Pressure: One to six channels
  • Oxygen: One to six channels • Auxiliary (optional): Two per installed channel; may be used for 0/4 to 20 mA and 0 to 4 V devices
Analog Outputs Two per channel; internally optically isolated; 10-bit (0.1%) resolution
Programmable Switch-Selectable Outputs
  • 0 to 2 V, 10 S minimum load resistance
  • 0/4 to 20 mA, 400 S maximum series resistance
Digital Outputs RS232 serial communications port; information is transmitted as ASCII
A/D Resolution 12-bit or 16-bit with MIS probe
Computer-Enhanced Response Software Standard: Provides fast measurements in abrupt dry-down applications (three- to five-minute speed of response)
Data Logging Up to 12 parameters simultaneously; stored in RAM; log length depends on number of items and time interval
Memory
  • Standard: Battery-backed 64 kB RAM
  • Optional: Additional 1 MB (or more) with PCMCIA card
PC-Interface Software Optional: PanaView™ instrument interface software
Alarm Relays
  • Two optional Form C relays per channel SPDT; rated for 2 A at 28 VAC/28 VDC
  • Standard and hermetically sealed (for Division 2 hazardous areas per CSA) available for high and low limits on each channel; set to any level within the range of the instrument
Alarm Set Point Accuracy ±0.1°C dew point
Display Graphic/text backlit LCD
Display Functions Displays up to six channel/parameter combinations simultaneously using text and/or graphics
Power Requirements Universal power supply adjusts automatically for 90 to 260 VAC, 50/60 Hz
Temperature
  • Operating: 32°F to 140°F (0°C to 60°C)
  • Storage: –22°F to 158°F (–30°C to 70°C)
Data Retention Calibration data stored in battery-backed RAM
Autocalibration Occurs on power-up and at user-selected time intervals
Configuration Dimensions (h x w x d)
  • Rack mount: 5.22 x 19.00 x 17.03 in (13.26 cm x 48.26 cm x 43.26 cm)
  • Bench mount: 5.85 in x 13.94 in x 17.03 in (14.86 cm x 35.41 cm x 43.26 cm)
  • Panel mount: 8.25 in x 16.50 in x 17.03 in (20.96 cm x 41.91 cm x 43.26 cm)
  • Weatherproof: 22.23 in x 16.23 in x 9.48 in (56.46 cm x 41.22 cm x 24.08 cm)
  • Explosion-proof: Consult factory

Moisture Measurement

Type GE Moisture Image Series, TF, and M Series thin-film aluminum oxide probes
Calibration Ranges (Dew/Frost Point)
  • Standard: 68°F to -112°F (20°C to –80°C) with data to -166°F (–110°C)
  • Ultralow: -58°F to -148°F (–50° to –100°C) with data to -166°F (–110°C)
  • Extended high: 140°F to -112°F (60°C to –80°C) with data to -166°F (–110°C)
Accuracy (Dew/Frost Point)
  • ±1.1°F (±2°C) from 140°F to -85°F (60°C to –65°C)
  • ±1.6°F (±3°C) from -87°F to -166°F (–66°C to –110°C)
Repeatability (Dew/Frost Point)
  • ±0.3°F (±0.5°C) from 140°F to -85°F (60°C to –65°C)
  • ±0.5°F (±1.0°C) from -87°F to -166°F (–66°C to –110°C)
Operating Pressure 5 of Hg to 5000 psig (345 bar) limited by optional pressure sensor—see pressure sensor ranges
Other Moisture Parameters Relative humidity (RH), parts per million by volume (ppmv), parts per million by weight (ppmw), parts per billion by volume (ppbv), pounds per million standard cubic feet (lb/MMSCF)

Temperature Measurement

Type Optional thermistor built into moisture probe
Range –22°F to 158°F (–30°C to 70°C)
Accuracy ±0.3°F (±0.5°C) at –22°F (–30°C)

Pressure Measurement

Type Optional transducer built into TF and Moisture Image Series moisture probes or standard external pressure transmitter, 24 V, 4 to 20 mA
Range 0 to 5000 psig (1 to 345 bar); specify type and pressure range
Accuracy ±1% of full scale
Pressure Rating Three times the span of the available range to a maximum of 7500 psig (518 bar)

Oxygen Measurement

Optional nondepleting electrolytic oxygen sensor.
Additional details and specifications are available on request.

This multichannel analyzer measures moisture in gases and non-aqueous liquids and oxygen (optional) in gases. Designed for permanent installations, it is used in conjunction with Moisture Image Series, TF and M Series moisture probes for industries including:

  • Petrochemical
  • Natural gas
  • Industrial gas
  • Semiconductor
  • Furnace gas/heat treating
  • Power generation
  • Air dryer
  • Pharmaceutical
  • Aerospace

Moisture Monitor Series 3 Analyzer (MMS 3)

The Moisture Monitor Series 3 Analyzer (MMS 3) analyzer is a single- or dual-channel analyzer with a two-line, alphanumeric LCD for moisture, oxygen, temperature and other parameters. It is available as a battery-powered portable with data logging, making it ideal for spot sampling and storing data from one or two locations without reprogramming. It also offers built-in intrinsic safety for all sensor inputs. The MMS 3 is available in rack, bench, panel, weatherproof and explosion-proof packages.

Electronics

Intrinsic Safety Galvanic isolation and energy-limiting circuitry for all inputs (except auxiliary):
BAS01ATEX7097
II (1) G [EEx ia] IIC –20°C to 50°C;
CSA C US Class I, Division 2, Groups B,C&D T4A–Assoc.
Elec. Apparatus [Exia] and Class I, Division 1,
Groups A, B, C & D T4
European Compliance Complies with EMC Directive 89/336/EEC and 73/23/EEC LVD (Installation Category II, Pollution Degree 2)
Inputs Moisture: One or two channels
  • Temperature: One or two channels
  • Pressure: One or two channels
  • Oxygen: One or two channels
  • Auxiliary (optional): Two per installed channel; may be used for 0/4 to 20 mA and 0 to 2 V devices
Analog Outputs Two per channel; internally optically isolated; 12-bit (.025% resolution)
Programmable Switch-Selectable Outputs
  • 0 to 2 V, 10 kΩ minimum load resistance
  • 0/4 to 20 mA, 400 Ω maximum series resistance
Digital Outputs RS232 serial communications port; information is transmitted as ASCII characters
Computer-Enhanced Response Optional: Software provides fast measurements in abrupt dry-down applications (three- to five-minute speed of response)
Data Logging Up to 12 parameters simultaneously; stored in batterybacked RAM (64 kB); log length depends on number of items and time interval
Memory
  • Standard: 64 kB RAM
  • Optional: Additional 1 MB or more with PCMCIA card
PC-Interface Software Optional: PanaView™ instrument interface software
Alarm Relays
  • Two optional Form C relays per channel SPDT; rated for 2 A at 28 VAC/28 VDC
  • Standard and hermetically sealed (for Division 2 hazardous areas per CSA) available for high and low limits on each channel; set to any level within the range of the instrument
Alarm Set Point Accuracy ±0.18°F (±0.1°C) dew point
Display 2 line x 20 character backlit LCD
Display Functions Displays two parameters
Power Requirements
  • Universal power supply adjusts automatically for 90 to 260 VAC, 50/60 Hz
  • Rechargeable battery pack (optional) lasts 8 hours under normal use and fully recharges in 16 hours
Temperature
  • Operating: 32°F to 140°F (0°C to 60°C)
  • Storage: –22°F to 158°F (–30°C to 70°C)
Configuration Dimensions (h x w x d)
  • Rack mount: 5.22 in x 19.00 in x 17.03 in (13.26 cm x 48.26 cm x 43.26 cm)
  • Bench mount: 6.35 in x 8.29 in x 17.03 in (16.13 cm x 21.06 cm x 43.26 cm)
  • Panel mount: 8.25 in x 10.85 in x 17.03 in (20.96 cm x 27.56 cm x 43.26 cm)
  • Weatherproof: 22.23 in x 16.23 in x 9.48 in (56.46 cm x 41.22 cm x 24.08 cm)
  • Explosion proof: Consult factory

Add 1.65 in (4.19 cm) to accommodate strain relief tabs

Moisture Measurement

Type GE TF and M Series thin-film aluminum oxide probes
Calibration Ranges
  • Standard: 68°F to -112°F (20°C to –80°C) with data to -166°F (–110°C)
  • Ultralow: -58°F to -148°F (–50°C to –100°C) with data to -166°F (-110°C)
  • Extended high: 110°F to -112°F (60°C to –80°C) with data to -166°F (–110°C)
Accuracy (Dew/Frost Point)
  • ±3.6°F (±2°C) from 140°F to -85°F (60°C to –65°C)
  • ±5.4°F (±3°C) from -86°F to -166°F (–66°C to –110°C)
Repeatability (Dew/Frost Point)
  • ±0.9°F (±0.5°C) from 140°F to -85°F (60°C to –65°C)
  • ±1.8°F (±1.0°C) from -86°F to -166°F (–66° C to –110°C)
Operating Pressure 5 μ of Hg to 5000 psig (345 bar) limited by optional pressure sensor—see pressure sensor ranges
Other Moisture Parameters Relative humidity (RH), parts per million by volume (ppmv), parts per million by weight (ppmw), parts per billion by volume (ppbv), pounds per million standard cubic feet (lb/MMSCF)

Temperature Measurement

Type Optional thermistor built into moisture probe
Range –22°F to 158°F (–30°C to 70°C)
Accuracy ±0.9°F (±0.5°C) at -22°F (–30°C)

Pressure Measurement

Type
  • Optional transducer built into moisture probe
  • Variety of standard GE external transmitters
  • Any user-supplied pressure transmitter, 24 V powered, 4 to 20 mA
Range 0 to 5000 psig (1 to 345 bar); specify type and pressure range
Accuracy ±1% of full scale
Pressure Rating Three times the span of the available range to a maximum of 7500 psig (518 bar)

Oxygen Measurement

Optional nondepleting electrolytic oxygen sensor. Additional details and specifications are available on request.

This Panametrics aluminum oxide moisture sensor probe measures moisture concentration in gases and non-aqueous liquids from trace to ambient levels. It is designed to be used in conjunction with all Panametrics moisture analyzers for industries including:

  • Petrochemical
  • Natural gas
  • Industrial gas
  • Semiconductor
  • Furnace gas/heat treating
  • Power generation
  • Air dryer
  • Pharmaceutical
  • Aerospace

Moisture Monitor Series 35 (MMS 35)

The Moisture Monitor Series 35 (MMS 35) single-channel hygrometer combines our technologically advanced M Series moisture probe with state-of-the-art electronics for superior speed of response, stability and overall performance.

Moisture Probe

Intrinsic Safety Intrinsically safe when connected to a Panametrics Moisture Series analyzer, PM880 portable hygrometer or intrinsically safe barriers in accordance with the user’s manual.
M Series moisture probe:
  • BAS01ATEX1096X
  • II 1 G EEx ia IIC T4 (-20°C to +80°C), and
  • CSA C US Class I, Division 1, Groups A,B,C&D T4,
  • LR44204-23
Optional Electrical Heater Complies with EMC Directive 89/336/EEC and PED 97/23/EC for DN<25
Type Aluminum oxide moisture sensor
Calibration Each probe is individually computer calibrated against known moisture concentrations, traceable to NIST or NPL.
Dew/Frost Point Calibration Ranges • Overall capability: 140°F to -166°F (60°C to –110°C) by request
• Standard calibration range: 68°F to -112°F (20°C to –80°C) with data to -166°F (-110°C)
• Ultralow calibration range: -58°F to -148°F (-50°C to –100°C) with data to -166°F (-110°C)
Accuracy • ±3.6°F (±2°C) in range of 140°F to -85°F (60°C to –65°C)
• ±5.4°F (±3°C) in range of -86°F to -166°F (-66°C to 110°C)
Repeatability • ±0.9°F (±0.5°C) in range of 140°F to -85°F (60°C to –65°C)
• ±1.8°F (±1.0°C) in range of -86°F to -166°F (–66°C to –110°C)
Temperature • Operating: -166°F to 158°F (–110°C to 70°C)
• Storage: Maximum 158°F (70°C)
Operating Pressure (Dependent on Mount) • Flange process connection (M1): 5 :Hg to 75 psig (6 bar)
• Threaded process connection (M2): 5 :Hg to 5000 psig (345 bar)
Flow Range • Gases: Static to 10,000 cm/s linear velocity at 1 atm
• Liquids: Static to 10 cm/s linear velocity at density of 1 g/cc
Response Time Less than five seconds for 63% step change in moisture content in either wet-up or dry-down cycle
Input Voltage 1 VAC, 77 Hz
Impedance Range 50 kS to 2 MS depending on water vapor
Limited Warranty • Calibration: Six months from delivery
• Materials and workmanship: One year from delivery

Optional Temperature Sensor

Type Nonlinear negative temperature coefficient (NTC) thermistor (resultant temperature linearized by microprocessor)
Operating Range –22°F to 158°F (–30°C to 70°C)
Accuracy ±0.9°F (±0.5°C) overall
Response Time (Maximum) 1 second in well-stirred oil or 10 seconds in still air for a 63% step change in increasing or decreasing
temperature
Optional Electrical Heater 120 VAC, 120 W
230 VAC, 75 W

This Panametrics aluminum oxide moisture sensor probe measures moisture concentration in gases and non-aqueous liquids from trace to ambient levels. It is designed to be used in conjunction with all Panametrics moisture analyzers for industries including:

  • Petrochemical
  • Natural gas
  • Industrial gas
  • Semiconductor
  • Furnace gas/heat treating
  • Power generation
  • Air dryer
  • Pharmaceutical
  • Aerospace

MTS6 Panel Mount OEM hygrometer

The MTS6 Panel Mount OEM hygrometer  with analog output and alarms.

Series 6 Specifications

  • Moisture Measurement
  • Sensor Type: Thin-film aluminum oxide
  • Moisture Probe Compatibility: Compatible with all GE M Series aluminum oxide moisture probes and VeriDri transmitters

Dew/Frost Point Temperature

  • Overall range capability: 140°F to –166°F (60°C to –110°C)
  • Standard 68°F to –112°F (20°C to –80°C) with data to -166°F (–110°C)
  • Extended 140°F to –85°F (60°C to –80°C) with data to 166°F (–110°C)

VeriDri ranges include:

  • –110°C to 20°C
  • –110°C to –50°C
  • –90°C to 10°C
  • –80°C to –30°C
  • –80°C to 20°C
  • –30°C to 20°C
  • –130°F to 70°F
  • –130°F to –40°F
  • –40°F to 70°F
  • –50°F to 50°F
  • –100°F to 0°F

Calibrated Accuracy at 77°F (25°C)

  • ±3.6°F (±2°C) from –85°F to 50°F (–65°C to 10°C)
  • ±5.4°F (±3°C) from –112°F to –86°F (–80°C to –66°C)

Repeatability

  • ±0.9°F (±0.5°C) from –85°F to 50°F (–65°C to 10°C) dew/frost point
  • ±1.8°F (±1.0°C) from –112°F to –86°F (–80°C to –66°C) dew/frost point


This single-channel analyzer measures moisture in compressed air and gases. Designed for permanent installations, it is used in conjunction with M Series moisture probe or VeriDri moisture transmitter for air/gas dryer original equipment manufacturer (OEM) applications:

  • Regenerative desiccant air or gas dryers
  • Plastic drying applications prior to injection/blowmolding
  • Medical air dryers
  • Glove box monitoring

MMY 2650

The MMY 2650 is a moisture analyzer designed to operate with the General Eastern DY 5 planar gold aluminum oxide dew point sensor. The unit offers an optional input that can be used for pressure compensation, an external temperature probe, or any other sensor this input can be configured to accept 0 to 5 VDC and 4 to 20mA. The standard product comes with one current output (0-20 mA or 4-20 mA), a system alarm relay, and 3 adjustable relays, all with single-pole double-throw 2.5A contacts, selectable failsafe mode, and adjustable hysteresis. One or two additional current outputs are available.

Keywords: analyzer, Moisture Measurement

Contact Veronics

Email - none provided

Print this page

Other products and services from Veronics

09-12-2011

Transmitters | Veronics Instruments Inc

The Veridri Moisture Transmitter provides Dryer Manufacturers and other OEM’s with the means to make simple accurate and reliable moisture measurements with a wide variety of Dewpoint ranges. The MMY30 is specifically designed to measure dew point or ppm at line pressure or atmospheric pressure for control of desiccant air dryers. The MMY31 mounts directly in-line for applications such as glove boxes, where a bypass is not appropriate.


09-12-2011

Portables | Veronics Instruments Inc

The PM880 system consists of the battery-powered moisture analyzer, moisture probes, battery charger, cables and accessories, and is packaged in a durable soft carrying case with a handle and shoulder strap. When purchased with the internal sample system option, a built-in moisture probe and sampling system with process connections are included, making it a completely self-contained hygrometer system.


09-12-2011

Handhelds | Veronics Instruments Inc

The Protimeter MMS (Moisture Measurement System) combines moisture meter and hygrometer functionality in one instrument to give full moisture measurement and dampness diagnostic capability. The Grainmaster i and the Grainmaster 900e meters enable the user to measure moisture and temperature levels in ground samples of a wide range of crops.


09-12-2011

Calibration | Veronics Instruments Inc - calibration

We produce our own very dry reference gas and use a primary standard, the Optica Chilled Mirror hygrometer, as our reference, the MG101 as a calibrated Moisture Generator and the Legacy Brand of Panametrics Microprocessor based hygrometers as a secondary reference and for recording of data. Accurate field calibration/verification of moisture sensors. Repeatable levels of water vapour in a carrier gas.


09-12-2011

Chilled Mirror | Veronics Instruments Inc

The Optica measures dew point temperature by regulating the temperature of a polished metal mirror by the use of optical feedback such that a constant mass of dew or frost is maintained. The temperature of the mirror is measured with a precise PRTD and is by definition equal to the dew or frost point. This Primary Standard has the following Features and Benefits.


09-12-2011

Services | Veronics Instruments Inc

On-Site Liquid Flow Studies – Veronics factory trained technicians can provide you with flow measurements using our Portable Non Intrusive PT878 Transit Time Clamp-On flowmeter for pipe sizes 1/2. Steam Trap Surveys – With the help of the U.E. Systems Ultrasonic Inspection System Veronics can perform Steam Trap Surveys at your location and take the guess work out of Steam Trap Maintenance.